Instructions for using the ATP meter :
Main features of ATP measuring instrument :
1. The ATP rapid detector adopts special sealing material to enhance the light-shielding property, and the detection result is more accurate and stable.
2. The interface is simple and easy to operate.
3. The bottom detection is not affected by the hand-held or placed angle of the instrument, the detection data is not disturbed, and the result is more stable.
4. Integrated box design, convenient for travel.
5. Having a significantly low background value is more conducive to the detection of trace ATP.
6. With good reproducibility, to provide users with more reliable and accurate data.
7. Metal case, anti-oxidation, strong aging, increase the service life.
ATP measuring instrument operation steps:
The swab test tube contains a unique high sensitivity liquid stable integrated reagent. The swab test tube detects the total ATP activity of bacteria or other microorganisms on the surface of the object and food residues, giving a quick and comprehensive clean test result.
Swab detection tube use precautions:
â—† After removing the swab test tube from the freezer, please leave it at room temperature for about 5 minutes;
â—†The swab test tube after sampling can be placed in the dark for about 4 hours;
â—†Do not touch the swab or swab with your hand;
â—† After the sample box solution in the swab test tube is reacted, it should be placed in the fluorometer and read within 20 seconds.
â—† The swab solution has been diluted to safely detect the surface of the food.
Note: Please keep it in the dark before use. If the surface of the extracted sample is dry, please wash the cotton swab head with pure water or distilled water to ensure more efficient extraction of the sample. The instrument should be tested vertically when testing, otherwise the test result is likely to be zero!
ATP measuring instrument operation steps:
1. Hold the tube, remove the wetted swab from the test tube, and apply the test area (about 4 cm * 4 cm).
2. Insert the swab into the test tube.
3. Hold the suction valve, pinch the tube with the other hand, and break it from the suction valve.
4. Lightly squeeze the tube twice, squeeze the reagent in the tube, shake it two or three times, and mix well with the swab after the test tube.
5. Place the entire test tube in the fluorometer, close the hatch cover, and measure.
Maintenance
a) daily maintenance
Daily clearing:
The outer casing of the instrument can be cleaned with a neutral detergent and the organic solvent is disabled.
It must be stopped before maintenance, please refer to the shutdown operation procedure.
b) Maintenance during operation
When the red paper on the printer's paper indicates that the paper is running out, replace the paper as soon as possible. After the replacement, the paper cover of the printer should be pressed down and installed.
c) maintenance cycle
After 12 months of use, the instrument should be routinely serviced to check for loose parts.
d) normal maintenance procedures
Touch screen maintenance:
Due to temperature, power supply voltage variation or other factors, the touch screen may not display properly or the display may be too pale or too dark. In this case, contact the relevant personnel for repair.
Moving parts maintenance:
After the instrument has been in operation for 12 months, you can contact a professional engineer to perform routine testing on the instrument.
e) Maintenance during long-term decommissioning
If the instrument is not used for a long time, after the instrument has been routinely cleaned, protect the instrument with a dust cover and store it in a dry and cool place.
Transportation and storage
a) Packaging and transportation precautions
The packaged analyzer should be transported according to the specified conditions. When transporting, it should be handled lightly to ensure the integrity of the outer box, pay attention to waterproof and moisture.
b) Storage conditions, storage period and precautions
Storage temperature of the instrument: -20 ° C ~ 55 ° C.
Storage humidity of the instrument: Relative humidity is less than 93% and there is no condensation.
Other requirements: no corrosive gases and well ventilated areas.
Common fault analysis and elimination
error code | Fault phenomenon | Cause Analysis | Method of exclusion |
(no) | No display at boot, no action on the instrument | The power adapter is not connected or is damaged | Reconnect the power adapter or contact the manufacturer |
Printer flashes | The printer light is blinking and the instrument cannot print the result | The printer is out of paper | Install new printer paper |
Instrument short tone | Non-stop short tone | The user took out the sample being read | Reload the sample |
CSY-ATP measuring instrument technical parameters:
☆ Detection accuracy: 1*10-15mole atp
☆ coliform group: 1-106cfu
☆Allergens: 1-100ppm
☆ background: less than 20RLU
☆ Detection range: 1-999999RLUs
☆ Detection time: 10 seconds ☆ Repeatability: CV% value (coefficient of variation) is not more than 7.4%
☆ Sampling point setting: no less than 2000 ☆ Storage function: no less than 10,000 test results ☆ Interface technology: USB
Digital Radiography made affordable, Created to meet the needs of community to hospitals and private radiology practices, it enables the price-sensitive customers to join the drive to go digital.The versatile floor mounted radiography system enables you to go from film to CR to DR. Flooe-mounted easy for installation and operation.
The United-imaging is floor mounted system comprises a radiographic table with integral floor guide rail and a wallstabd. Requiring little room preparation, it is easy to install.
United-imaging according to the international advanced processing mode and standardlize design. the parts are made with precision CNC machine and moulding. Which adopt high degree of standardized, reasonable and compact structure. With reliable, durable and elegant appearance and advanced processing technology.
X-Ray Digital Machine,X-Ray Digital Machine In Medical,X-Rays Digital Medical Machine,Medical X-Ray Digital Imaging Machine
Shanghai Rocatti Biotechnology Co.,Ltd , https://www.shljdmedical.com