Dust particle counter is mainly used to test what

Dust particle counter is mainly used to test what

Dust particle counter is mainly used to test what

CLJ-E type (semiconductor laser) dust particle counter (LED display)

The technical indicators of this series of instruments meet the requirements of the JJF1190-2008 verification regulations promulgated by the State Administration of Metrology. The whole machine functions adopt American micro-computer control processing technology and internationally advanced SMT chip chip packaging technology and semiconductor laser sensor technology and anti-noise air pump. The test results can be printed directly. It has many functions, high measurement accuracy, fast speed, easy to carry and easy to operate. The instrument can simultaneously measure the number of dust particles of various particle sizes at one time, and can choose to observe the number of particles of a certain particle size and its changes, which is very convenient for researching, testing and evaluating various clean environments. This series of instruments has advanced performance design and stable and reliable quality.
This series of products has been widely used in clean room testing; filter on-site inspection, leak detection; can monitor ultra-clean workbench, biosafety cabinet, HVAC system, computer room, beverage packaging environment, medical device production environment, hospital clean operating room, Automotive spraying environment Microelectronics, pharmaceutical, biochemical products, food hygiene, fine chemicals, precision machinery and aerospace production and scientific research departments, is the pharmaceutical companies and their supervision and management departments to implement GMP standards and electronic production enterprises.
Main technical parameters and performance:

1. Light source : all semiconductor laser source, imported diode laser

2. Sample size: 2.83L/min (0.1cfm/min)

3. Detection range: 100 to 300,000

4. Allow the dust concentration of the tested air to be 100,000 / 2.83L

5. Particle size channel: 0.3 0.5 1.0 3.0 5.0 10.0 (μm) six files
6. Display or print can convert the particles contained in 2.83 liters/min into the number of particles contained in 1 m 3 .

7. Sampling period: 1~10 (min)

8. Self-cleaning time: ≤15 (min)

9. Calibration: It can be traced back to the National Institute of Standards and Technology (NIST). Our company has passed the national measurement and construction assessment, which can be traced back to the Shanghai Institute of Metrology and Measurement Technology. It can be self-calibrated or calibrated by a third-party national metrology institute.

10. Working environment: Temperature : 10~35°C Relative temperature : 20~75%RH

11. Power consumption: 25W

12. Measure the range and accuracy of temperature and humidity: (optional)

(1) Temperature: 0 to 50 ° C ± 1 ° C

(2) Humidity: 0 to 100% RH ± 5%

13. Number of sampling points 2 to 7 points setting

14. Number of sampling points per point 2 to 9 times setting

15. UCL report: in line with FS-209E, China GMP standards

16. Working hours: 8 hours

17. Power supply: AC220V±10% 50±2Hz

18. Weight : 4.8kg

19. Dimensions: 260 × 130 × 340

20. Six-stage particle size dust concentration is detected simultaneously, followed by digital display or self-selected particle size display.

21. Remarks: Built-in printer, automatic judgment of purification level, power sampling head, sampling rack